GAS ANALYSIS OXYMET
The analyzers OXYmet are based on a miniaturized oxide cell, using the current limiting effect due to Knudsendiffusion. Thank to this innovative technology, the OXYmet analyzers offer excellent performances concerning accuracy, long term stability and long life.
Functions
The basic functions satisfy a lot of applications in industrial, food and medical environments, are include:
- Microcontroller management
- 4 digit 7 segments display
- 2 set-point with relay outputs, fully configurable by user
- Acoustic alarm
- 4...20mA analog output, linearized
- Serial output RS 232 (standard) / RS 485 (optional) ModBus protocol RTU
- Compatible cell for flow analysis and environmental monitoring
- Zero and Span calibration adjustable by appropriate commands on your device menu
Applications
Process monitoring for synthetic air in hospital
Process monitoring for air + O2 in hyperbaric chambers
Air monitoring for packaging food applications - Process monitoring for industrial and laboratories
Technical data
Power supply 220 Vac 50 / 60 Hz, P max 6 VA, fuse 0,5 A
Measuring range 0...2,5 %, 0...25 %, 0...70 %, 0...95% Vol O2
Visual indication N°4 digit 7 segments display
Allarme Set Alarm set - point 1: red LED Ø 5 mm
Allarme Set Alarm set - point 2: red LED Ø 5 mm
Speaker max 95 dB @ 1 meter
Max. load 700 Ohm open voltage > 15 Vdc negative common configuration
Relay outputs Vmax = 60 Vdc / 220 Vac / Imax = 1A
N.O. and N.C., output switch max. voltage 60 Vdc / 220 Vac; max. current 1A
Response time < 30 sec. From 0 to 99 % F.S.
Warm - up time 4 minutes
Combined error < ± 1% sul F.S. - full range
Atmosperic pressure drift Irrelevant (included in combined error)
Long term drift < 2% sul F.S. in 5 anni - < 2% of F.S. over 5 years
Flow range 0,5 - 15 l/m (only with flow applications)
Pneumatic connections 1/8" F
Housing Black plastic box type DIN 43700, panel mounting, dim. 96 x 96 x 169 mm
Serial output bidirezionale - opzionale RS 232 undirectional - standard, RS 485 bidirectional (optional)
Weight 750 gr.
Operating temperature 0...60°C
Reference standars and directives CE